Portable Ion Chromatograph CLPIC-601 is equipped with a dual-piston serial pump to deliver exceptional precision. It offers maximum pressure of 35 MPa to enable efficient ion separation. Incorporated with a user-friendly interface for seamless and hassle-free operation. Designed for portability and ease of transport to field locations. Our Portable Ion Chromatograph ensures consistent accuracy and reliability, even in challenging environments.
Pump Type | Dual-Piston Serial Pump |
Pump Maximum Pressure | 35 MPa |
Pump Flow Rate Range | 0.001 to 9.999 mL/min |
Pump Flow Stability | ≤ 0.67% |
Tube Diameter | 1/16 |
Valve Rotor Material | PEEK |
Valve Medium Material | PEEK |
Column Heater Temperature Range | Ambient + 5 to 60 ℃ |
Heater Temperature Accuracy | ± 0.2℃ |
Temperature Stability | ≤ 0.2℃/h |
Columns Supported | 2, 3, 4, and 5 mm |
Analytical Column Length | 250 mm |
Guard Column Length | 50 mm |
Conductivity Detector Type | Auto-Range Conductivity Detector |
Cell Volume | ≤ 0.8 µL |
Detection Range | 0 to 45000 µS/cm |
Resolution | ≤ 0.0020 µS/cm |
Output Voltage | -6000 to + 6000 mV (Adjustable) |
Baseline Noise | ≤ 0.0028 µS/cm |
Baseline Drift | ≤ 0.0293 µS |
Operating Temperature Range | Ambient + 5°C to 60°C |
Detector Temperature Accuracy | ± 0.01℃ |
Temperature Compensation | 1.7% / ℃ |
Maximum Pressure | 10.0 MPa |
Instrument Linearity | ≥ 0.999 |
Quantitative Repeatability | ≤ 0.71% |
Qualitative Repeatability | ≤ 0.07% |
Minimum Detectable Concentration | Cl⁻: ≤ 0.0012 μg/mL |
Suppressor Type | Self-Regeneration Electrolytic Micro-Membrane Suppressor |
Suppressor Maximum Pressure | 6.0 MPa |
Non-Suppressed Chromatography | Yes/Support |
Void Volume | <40 μL (4 mm), <15 μL (2 mm) |
Power Supply | 220V AC |
Battery Type | Built-in Lithium Battery |
Dimension | 420 × 320 × 190 mm |
Net Weight | ≤ 9 kg |
Gross Weight | 13 kg |
Connectivity
Leak Protection
External Power Supply
Auto-Range Detector
Electrolytic Micro-Membrane Suppressor
Portable Eluent Generator
Portable Ion Chromatograph CLPIC-601 is ideal for environmental monitoring, industrial effluent analysis, on-site quality control, real-time testing, manufacturing processes, field-based experiments and many more.
High-Pressure Advection Pump
Conductivity Detector
Manual 6-Way Valve
Anion Column
Guard Column (Anion)
Anion Suppressor
Chromatography Software
Column Heater
Spare Parts Kits
Vacuum Degassing Pump
Sand Core Filtration Device
0.22um Syringe Filter × 100 Pieces
Power Supply
Ipad
Cation column
Guard column (cation)
Cation suppressor
Ask Our Experts
Frequently Asked Questions
Its compact design, lightweight construction, and user-friendly interface make it highly portable and easy to operate in field conditions without sacrificing performance.
The system features a dual-piston serial pump, which ensures stable flow and high precision during analysis—perfect for reliable results in any location.
Yes, our model supports high-pressure operations, enabling effective ion separation and stable performance across various samples and environments.
It supports a wide range of column sizes, allowing flexibility in sample analysis and compatibility with different separation requirements.
Yes, it supports both suppressed and non-suppressed chromatography, giving users versatile options depending on sample type and analytical goals.
It includes a precise column heater and detector temperature control, which ensures consistent thermal conditions for accurate and repeatable results.
Commonly detected ions include anions like fluoride, chloride, nitrate, sulfate, and cations like sodium, potassium, calcium, and magnesium.
Our ICs are equipped with suppression technology to improve sensitivity and reduce background noise.
Water, soil extracts, wastewater, beverages, and other aqueous samples can be analyzed, often with minimal sample prep.
Yes, calibration with standard solutions is typically done before field measurements to ensure accuracy.
They are used in environmental monitoring, water quality testing, industrial process control, field research, and emergency response situations.