Liquid Chromatograph Mass Spectrometer CLMS-602 provides a mass range of 10 to 2000 amu for precise detection. It supports an HPLC flow range of 0.001 to 9.999 mL/min for diverse analytical applications. Designed for fast data acquisition to enable real-time and targeted analysis. Equipped with ESI and APCI ionization modes to enhance detection sensitivity. Our LCMS ensures consistent and accurate performance with minimal manual adjustments.
Mass Range | 10 to 2000 amu |
Resolution | FWHM ≤ 0.6 amu |
Scan Speed | 1000 amu/s (standard), 10000 amu/s (fast mode) |
Mass Axial Stability | ±0.2 amu per 12 hrs |
Peak Area Repeatability | RSD ≤ 4.6% |
Retention Time Repeatability | RSD ≤ 1.5% |
Polarity Switching Speed | 20 ms |
Ion Source Flow Rate | ESI: 11 to 1000 µL/min APCI: 50 to 2000 µL/min |
Ion Source Temperature | 40 to 200℃ |
Corona Current (APCI) | 0 to 10 µA |
HPLC Flow Range | 0.001 to 9.999 mL/min |
HPLC Flow Accuracy | RSD ≤ 0.06% |
HPLC Pressure Resistance | HPLC: Max 42 MPa |
Column Oven Temperature Range | Room temperature to 80℃ |
Working Temperature | 20℃ to 30℃ |
Humidity | 70% |
Power Supply | 220V ±10%, 10A, 50Hz |
Weight | 23 kg |
Advanced Mass Analysers
Selectivity
High-Resolution
Fast Analysis
Automated Calibration
Liquid Chromatograph Mass Spectrometer CLMS-602 is widely used in pharmaceuticals, environmental analysis, food safety, and clinical research, as well as in biotechnology, metabolomics, petrochemical testing, drug discovery, toxicology, and quality control laboratories.
Ask Our Experts
Frequently Asked Questions
Our LC-MS is engineered for precise mass spectrometry analysis, suitable for applications in pharmaceuticals, environmental testing, food safety, and chemical research.
It supports a wide HPLC flow range, making it compatible with various analytical methods and sample types. This allows flexibility across multiple scientific and industrial workflows.
The system is equipped with both Electrospray Ionization (ESI) and Atmospheric Pressure Chemical Ionization (APCI) modes, enhancing its ability to detect a broad spectrum of compounds with high sensitivity.
Our model offers excellent mass stability over extended durations, ensuring consistent and reliable performance throughout the day with minimal manual adjustments.
It ensures strong repeatability for both retention time and peak area, contributing to reliable and reproducible test results
Yes, the system is built to support high-pressure HPLC operation, ensuring durability and compatibility with demanding analytical methods.
It is designed to operate efficiently in standard laboratory environments with stable room temperatures and moderate humidity levels.
LC-MS stands for Liquid Chromatography-Mass Spectrometry. It combines the separation power of liquid chromatography with the identification capabilities of mass spectrometry to analyze complex mixtures.
LC-MS is widely used in pharmaceutical research, clinical diagnostics, environmental testing, food safety, toxicology, metabolomics, proteomics, and biotechnology for detecting and measuring a wide range of substances.
Our LC-MS can be used for both qualitative identification and quantitative measurement of analytes, often with very high precision and accuracy.
Yes. Our system features autosamplers, software-driven workflows, and automated calibration, allowing for high-throughput and unattended operation.
The analysis time ranges from 5 to 45 minutes, depending on the method, sample type, and complexity.
Mass Range: m/z 2 to 2,000
Resolution: R < 0.7 u FWHM and adjustable to 0.5 u